Abstract

An example of uncertainty evaluation of an efficiency curve is reported. In the applied method, the efficiency curve is determined through the weighted least square method based on measured efficiencies employing a mixed-radionuclide standard source. The variance-covariance matrix of the efficiencies is determined by evaluating the uncertainties of the counting, the calibrated radioactivities, the correction factors for true coincidence summing, and the inadequacy of the fitting function. The proposed method can be a basis to develop a more general procedure.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.