A portable target controlled infusion system for propofol has been developed based on a Psion hand-held microcomputer and the Ohmeda 9000 syringe pump. The system uses a pharmacokinetic model which describes the distribution and elimination of propofol to achieve and maintain any selected target blood concentration. Target blood concentrations of 1 micrograms/ml, 3 micrograms/ml and 5 micrograms/ml were selected in laboratory trials and the cumulative volumes delivered by the Psion system each minute were compared with the theoretical output calculated by the pharmacokinetic model. The results obtained showed that the computer system delivered volumes which were always within 2% of the theoretical values. This system offers a convenient and simple method of maintaining anaesthesia using propofol.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.