Abstract
In vivo dose verification in brachytherapy requires a small insertable dosimeter with a real-time readout capability. Fibre optic scintillation dosimeters, consisting of a plastic scintillator coupled to an optical fibre, are one of the most promising dosimeters for this application. We have developed two sizes of the BrachyFOD™ scintillation dosimeter which have external diameters of 2.2 mm and 1 mm and have determined their important dosimetric characteristics (depth dose relation, angular dependence, temperature dependence, energy dependence). We have shown that the background signal created by Cerenkov and fibre fluorescence does not significantly affect the performance in most clinical geometries using an 192Ir source from an HDR brachytherapy unit. The dosimeter design enables readout at less than 0.5 s intervals. The BrachyFOD™ satisfies the need for a real-time in vivo brachytherapy dosimeter.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.