Abstract
Accurate measurement of the permittivity and loss tangent of low-loss materials is essential due to their special applications in the field of ultra large scale integrated circuits and microwave devices. In this study, we developed a novel strategy that can accurately detect the permittivity and loss tangent of low-loss materials based on a cylindrical resonant cavity supporting the TE111 mode in X band (8–12 GHz). Based on an electromagnetic field simulation calculation of the cylindrical resonator, permittivity is precisely retrieved by exploring and analyzing the perturbation of the coupling hole and sample size on the cutoff wavenumber. A more precise approach to measuring the loss tangent of samples with various thicknesses has been proposed. The test results of the standard samples verify that this method can accurately measure the dielectric properties of samples that have smaller sizes than the high Q cylindrical cavity method.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.