Abstract

In this study, dispersive liquid-liquid microextraction based on the solidification of floating organic droplets was used for the preconcentration and determination of thorium in the water samples. In this method, acetone and 1-undecanol were used as disperser and extraction solvents, respectively, and the ligand 1-(2-thenoyl)-3,3,3-trifluoracetone reagent (TTA) and Aliquat 336 was used as a chelating agent and an ion-paring reagent, for the extraction of thorium, respectively. Inductively coupled plasma-optical emission spectrometry was applied for the quantitation of the analyte after preconcentration. The effect of various factors, such as the extraction and disperser solvent, sample pH, concentration of TTA and concentration of aliquat336 were investigated. Under the optimum conditions, the calibration graph was linear within the thorium content range of 1.0-250 µg L-1 with a detection limit of 0.2 µg L-1. The method was also successfully applied for the determination of thorium in the different water samples.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.