A simple and inexpensive method was developed with high performance liquid chromatography with PDA detection for determination of Adapalene and related impurities ((3-Adamantyl-4-methoxy) phenyl boronic acid, 6-(toluene-4-sulfonyloxy)-naphtalene-2-carboxylic acid methyl ester, 1-adamantyl-2-methoxybenzene, 6-(3-adamantan-1-yl-4-methoxy-phenyl)-naphtalene-2-carboxylic acid methyl ester and 3, 3’-di-(1-adamantyl)-4, 4’-dimethoxy-1, 1’-biphenyl). The chromatographic separations were achieved on (250×4.6 mm), 5.0 μm make: Phenomenex Luna column employing methanol, 0.1% orthophosphoric acid buffer, and Tetrahydrofuran in the ratio of 55: 30: 15 as mobile phase with gradient programmed at flow rate 1.0 mL/min was chosen. Five impurities were eluted within 35 minutes. The column temperature was maintained at 25°C and a detector wavelength of 260 nm was employed. The method was successfully validated by establishing System Suitability, Specificity, Linearity, Accuracy, limit of detection and Limit of quantification.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.