Abstract

Abstract Persistent episodes of hyperglycaemia in diabetes leads to serious damage to nerve system (referred to as diabetic neuropathy), such as nerve conduction abnormalities, pain and allodynia, and impaired axonal regeneration. Real-time monitoring glucose in brain is of great significance in better understanding the pathophysiological mechanisms of hyperglycaemia related brain disorders. However, it has never been achieved in real-time in vivo detection of the rise of glucose concentration in cerebrospinal fluid (CSF) caused by an instant glucose intake, due to technical difficulties. Here, we introduced a novel electrochemical approach with glass needle enclosed movable glucose probe to in vivo detect glucose in cisternal CSF of mice. After the glass needle was punctured into cisterna magna of mouse by stereotactic guidance, the inner probe was inserted into the cisternal CSF to achieve the recording. By this approach, we successfully measured the amperometric signal corresponding to the intraperitoneal injection of glucose solution. Thus, a novel approach was developed to real-time detect the glucose in cisternal CSF of mice in vivo.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.