Abstract

To study the effect on outflow facility and outflow resistance of a nitinol microstent implanted into Schlemm's canal. Using a constant pressure perfusion method, outflow facility and outflow resistance were measured in 26 pairs of dissected anterior segments from donated human eyes. Measurements were made at perfusion pressures of 10, 20, 30 and 40 mm Hg. The Hydrus Microstent was placed in Schlemm's canal of one eye and the contralateral eye underwent a sham procedure. Outflow facility and outflow resistance were measured again after the microstent implantation or sham procedure. The Hydrus Microstent significantly increased outflow facility from 0.33 ± 0.17 μL/min/mm Hg to 0.52 ± 0.19 μL/min/mm Hg (P < 0.001). Outflow resistance was significantly reduced from 4.38 ± 3.03 mm Hg/μL/min at baseline to 2.34 ± 1.04 mm Hg/μL/min (P < 0.001) with the microstent. There was a linear correlation between outflow resistance at baseline and decrease in outflow resistance with the microstent (R(2) = 0.89, P < 0.0001). The increase in outflow facility and decrease in resistance supports the potential use of the Hydrus Microstent as a surgical option to reduce intraocular pressure (IOP). The IOP-lowering effect may be higher in eyes with higher outflow resistance (and IOP) as compared with eyes with lower outflow resistance (and IOP).

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.