Abstract

A silicon-on-insulator diode array with a sensitive depth of 10 microns has been developed for microdosimetry in proton therapy. The detector was coupled to a radiation-hard charge sensitive amplifier with the probe assembly capable of measuring an LET down to 1.2 keV//spl mu/m. The device has been successfully tested at two proton therapy centers. The 230 MeV Northeastern Proton Therapy Center, Boston and the 250 MeV Proton Medical Research Center at Tsukuba, Japan. The device offers much improved spatial resolution compared with a proportional gas counter particularly in the critical high dose region around the proton Bragg peak. Due to its small cross-sectional area (0.04 cm/sup 2/) measurements may also be made in facilities with short high intensity beams.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.