Abstract
In this paper, we present a new optical computed tomography (Optical-CT) scanner for the verification of the radiation dose schemes delivered in modern radiotherapy applications. The optical-CT scanner is capable of providing rapid relative 3-D dosimetry with high spatial resolution with the use of normoxic N-Vinylpyrrolidone based polymer gel dosimeter. The scanner employs a diffuse uncollimated light illumination beam, a computer controlled motorized rotation stage and a charge-coupled device (CCD) camera. Various test experiments were performed to determine the performance characteristics of the optical-CT apparatus. Attenuation coefficient (micro) versus dose calibration data were generated from two calibration experiments using gel containers of two different diameters. All irradiations were performed using a 6 MV linear accelerator. A comparison of the reconstructed images between optical-CT scans using refractive index (RI) matching fluid and corresponding scans performed in free space was demonstrated. The dose readout of a test irradiation model was found to be in good agreement with independent readout performed by MR imaging. The findings presented in this study suggest that polymer dosimeters combined with the new optical-CT scanner constitute a potentially feasible method capable of measuring complex 3-D dose distributions with high resolution and in a wide dose range.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.