Abstract

Objective Assess the ability of a new device based on electrochemical principles using iontophoresis (the EZSCAN ®) to detect impaired glucose tolerance (IGT) and type 2 diabetes mellitus (DM). Methods Eligible Asian Indian subjects, n = 212, had anthropometric and blood pressure measurements, followed by an OGTT, HbA1c, serum lipids tests and EZSCAN ® measurement. Results Biochemically, 24 subjects were diagnosed with DM, 30 with IGT, 57 subjects had normal glucose tolerance (NGT) with metabolic syndrome (MS) and 101 had NGT without MS. Fasting plasma glucose (FPG) and HbA1c levels were highest in the DM group ( p < 0.0001 for both). HDL-C levels were different ( p = 0.015). FPG at a cut-off level of 7.0 mmol/L had a low sensitivity to detect DM (29%) EZSCAN ® had a 75% sensitivity to detect DM, 70% for IGT and 84% for NGT with MS at threshold >50%. Conclusions FPG had low sensitivity to detect DM in the study group. EZSCAN ® demonstrated good sensitivity to detect IGT and DM and also identified NGT with MS. The concept of measuring ion fluxes through the skin appears to be a powerful method for early detection of MS, IGT and DM.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.