Abstract

In this work, an improved method of ionic liquid based-single drop microextraction (IL-SDME) preconcentration was firstly coupled with tungsten coil electrothermal atomic absorption spectrometry (W-coil ET-AAS) detection. The new coupling was developed and applied for the determination of ultra-trace cadmium. Experimental parameters influencing the performance of IL-SDME as well as instrumental conditions were studied systematically, including IL-drop volume, chelating agent concentration, pH, stirring rate and time, heating program of W-coil ET-AAS, flow rate of carrier gas. Under the optimal conditions, the limit of detection (LOD) for cadmium was 0.015μgL−1 (sampling amount on W-coil was 10μL). The sensitivity enhancement factor was 42, while the improvement factor of LOD was 33. The established method was applied to determine cadmium in standard reference materials of rice and real water samples successfully. The developed IL-SDME-W-coil ET-AAS coupling represents a simple, green and highly sensitive method for cadmium determination.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.