Abstract

The aim of this study was to find out if Cobe Trima, a blood cell separator that automatically collects RBC, PLT and plasma, is adequate for routine multiple blood donation by apheresis. Eighty donors underwent multiple blood component donations by Cobe Trima. Blood counts were determined on the apheresis products to analyze their quality. Eighty procedures were performed collecting 193 products. The average platelet yield was 3.5x10(11) (+/- 0.46) in the 54 single product (SP) procedures and 7x10(11) (+/- 0.88) in the 26 double product (DP) procedures. WBC contamination of the PLT products was 1.7 x 10(5) (1.2-4.2). The mean platelet efficiency was 60 +/- 8.35% for SP and 66 +/- 9.59% for DP. The hemoglobin (Hb) content per unit was 46.21 g (+/- 7.84) in 8 DP and 40.82 g (+/- 6.41) in 34 SP procedures. The production of standardized blood components with good PLT yield and low WBC contamination plus high efficiency makes Trima one of the best blood cell separators of the new generation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.