Abstract
A green and efficient analytical approach was reported for simultaneous preconcentration, and separation of Pb(II) and Cd(II) in water, vegetables, and barbecue samples by dispersive solid-phase microextraction prior to their determination using flame atomic absorption spectrometry. A new poly-3-hydroxy butyrate-polyvinyl triethyl ammonium chloride comb-type amphiphilic cationic block copolymer (PHBvbNCl) was synthesized and characterized. Main variables such as pH, sorbent amount, adsorption time, eluent type, desorption time, and sample volume were optimized. Detection limits and working ranges for Pb(II) and Cd(II) were 0.03 μg L-1, 0.15 μg L-1, 0.1–250 μg L-1 and 0.5–375 μg L-1, respectively. Enhancement factor for Pb (II) and Cd (II) were 114 and 98. The adsorption capacity of PHBvbNCl for Pb(II) and Cd(II) was 175.2 mg g−1 and 152.9 mg g−1. After the accuracy of the method was confirmed by the analysis of certified reference materials, it was successfully applied to real samples. Finally, the analytical performance of the present method was compared with other methods.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.