Abstract

Bone marrow scintigraphy using radiocolloids labelled with 99Tcm allows imaging of the active marrow with moderate effort. However, most of the administered colloid is trapped in the liver and spleen giving rise to a considerable amount of scattered photons, which cause considerable problems when evaluating the adjacent bone marrow. There are scatter correction methods developed for single photon emission computed tomography. The application of such a technique to bone marrow scintigraphy in the liver region using two opposing views is described. Phantom measurement and patients studies show that substantially improved visualization of the bone marrow close to the liver is possible using this technique for scatter correction.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.