Abstract

A coincidence timing spectrometer was assembled using NaI(Tl) inorganic and BC418 type organic (plastic) scintillation detectors. The constant fraction timing method was used. Coincidence time resolution values of the detectors, which have great importance in TOF/PET measurements, were obtained separately. The resolutions were enhanced using a different method in a start-stop coincidence spectrometer: the signals from the start detectors were delayed. The results from the FLUKA Monte Carlo simulation code agreed well with the experimental results.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.