Abstract
The effect of metal artifact reduction algorithms on CT image quality was considered. The mean signal intensities and noise levels were compared for objects with different X-ray densities and spatial positions using phantom modeling. It was shown that, despite an improvement in the visual characteristics of the images, their quantitative characteristics may deteriorate: the change in the mean density reached in some cases 15%; the increase in noise, 30%.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.