Abstract

We present a low-temperature ultrahigh vacuum (UHV) scanning tunneling microscope setup with a combination of a superconducting solenoid coil and two split-pair magnets, providing a rotatable magnetic field up to 500 mT applicable in all spatial directions. An absolute field maximum of B=7 T(3 T) can be applied perpendicular (parallel) to the sample surface. The instrument is operated at a temperature of 4.8 K. Topographic and spectroscopic measurements on tungsten carbide and indium antimonide revealed a z-noise of 300 fm(pp), which barely changes in magnetic field. The microscope is equipped with a tip exchange mechanism and a lateral sample positioning stage, which allows exact positioning of the tip with an accuracy of 5 microm prior to the measurement. Additional contacts to the sample holder allow, e.g., the application of an additional gate voltage. The UHV part of the system contains versatile possibilities of in situ sample and tip preparation as well as low-energy electron diffraction and Auger analysis.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.