Abstract

A prototype valveless ventilator was attached by open deadspace tubing to a circle system and used to ventilate the lungs of 12 patients with low flows of anaesthetic gases for periods between 60 and 120 minutes during intra-abdominal surgery. Anaesthesia was induced with thiopentone and maintained with nitrous oxide 50% in oxygen and enflurane. This was reduced to 2 litres/minute after a 10-minute period of nitrogen wash out and stabilisation of anaesthetic gas concentration, with an initial anaesthetic gas flow of 6 litres/minute. The concentration of oxygen, carbon dioxide, nitrous oxide and enflurane were measured in the outflow from both the anaesthetic machine and the inspiratory limb of the circle system. The measured mean inspired oxygen and nitrous oxide concentrations showed no significant variation throughout the low flow period of the study. This new low flow open circle ventilation system appears to offer some advantages in terms of safety and versatility over other systems which are discussed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.