Abstract

Circular position sensitive detectors with sensitive areas of 17 cm 2 have been constructed by connecting gold strips on the front surface of a Li-drifted silicon detector to an external resistor network. Detector responses to alpha-particle sources and pulsed-light sources are presented which demonstrate the position and energy response characteristics of the detectors. Measurements with a pulsed-light source equivalent of 140 MeV show that a resolution (fwhm) of 0.19 mm (the width of the gold strips) can be achieved for energy deposits of ∼ 40 MeV. The position resolution (fwhm) obtained from these detectors mounted in a telescope exposed to 95–350 MeV/nucleon 40Ar nuclei is 0.82 mm for a single detector.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.