Abstract
This paper proposes a high frame rate analog front-end (AFE) IC combined with a piezoelectric micromachined ultrasonic transducer (PMUT) for ultrasound imaging systems (UISs). The proposed AFE IC, which consists of 8-channel transceivers and a peripheral circuit, employs analog multi-line acquisition (AMLA) based on sub-array beamforming. This AMLA generates multiple scanlines during a scan time by simultaneously and precisely controlling the delay of the AFE IC, thus achieving the high frame rate of the AFE IC. The sensitivities of the PMUT for transmitting and receiving ultrasound waves are simulated to be 67.5 kPa/V and 1.2 mV/kPa, respectively, which are high enough to mitigate the needs of the high-voltage electronics for UISs. The proposed AFE IC was fabricated using a 0.18-μm silicon-on-insulator process and occupies an active area of 2.56 mm × 1.68 mm. The measurement results using the simulated PMUT parameters show that the proposed AFE IC with the AMLA precisely controls its delay up to 1 μs at a clock frequency of 40 MHz. It also acquires more than 9,000 scan/s even at a focal depth of 150 mm, which is at least twice greater than those of prior works, achieving a high frame rate of 93 Hz when 100 scanlines are used for a frame. In addition, a figure-of-merit, which considers the energy consumption per channel/scan and the energy loss, is achieved to be 20 nJ/scan/dB at a focal depth of 150 mm and a center frequency of 3.6 MHz; this is the best result compared to previously reported works. Therefore, the proposed AFE IC is suitable for UISs requiring a high frame rate.
Highlights
Ultrasound imaging systems (UISs) have been widely researched for medical applications due to their advantages of non-invasiveness to the human body and real-time imaging capabilities
Multi-line acquisition (MLA) using a parallel processing technique based on plane wave imaging [2]-[6] was reported to increase the frame rate by producing multiple scanlines from transmitted unfocused ultrasound waves at a pulse repetition frequency (PRF)
ANALOG MULTI-LINE ACQUISITION The proposed analog multiline acquisition (AMLA) employs sub-array beamforming and produces multiple adjacent scanlines based on echo from the human body as transmitting the single-focused ultrasound beam, increasing the frame rate of the UIS
Summary
Ultrasound imaging systems (UISs) have been widely researched for medical applications due to their advantages of non-invasiveness to the human body and real-time imaging capabilities. This MLT caused a large dynamic power in the ultrasound probe during the multiple transmissions of the ultrasound waves and limited the SNR due to crosstalk between the simultaneously transmitted beams [10]. The AFE IC transmits high-voltage (HV) signals to the ultrasound transducers and receives echo from the human body These multi-line implementations in the digital domain have the advantages of generating multiple scanlines and allowing for flexible control of the beam formation.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.