Abstract

This paper presents an error compensation method for heartbeat intervals measured by a photoplethysmography (PPG) sensor. The objective of this work is to improve the accuracy of heart rate variability analysis (HRVA) using PPG, because a peak-to-peak interval (PPI) of the PPG has an error when compared with R to R interval (RRI) of electrocardiograph (ECG). This error is caused by several factors such as blood pressure fluctuation, body motion artifact, and body position. The proposed method predicts RRI using multiple regression analysis with time series of PPIs. The predicted RRI using the proposed method can also reduce the error of frequency characteristics, which are used for HRVA. Evaluation results with three subjects show that the proposed method achieved about 30% mean absolute error (MAE) reduction for predicted RRI and 17% mean absolute percentage error (MAPE) of LF/HF index of HRV analysis.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.