Abstract

A new and fully automated system with the interconnection of an Optical Immersion Probe (OIP) – pH meter – peristaltic pump was used to study the spectral and protolytic properties of carbocyanine the dyes 1,1′,3,3,3′,3′-hexamethylindocarbocyanine chloride (HIC); 1,1′,3,3,3′,3′-hexamethylindodicarbocyanine iodide (HIDC); and 3,3′-diethyloxadicarbocyanine iodide (DODC). This system can measure a large number of experimental points in a short time period. The effect of 32 various organic solvents on the UV-ViS spectra of the dyes was studied. The solvatochromic behaviour of studied dyes was characterized by positive solvatochromism for HIDC and negative solvatochromism for HIC and DODC. Through the application of a large number of experimental points, the protonation and hydrolysis constants of dyes were determined with high precision, where the confidence interval of the рK values is ±(0.001–0.005), compared with a confidence interval of ±(0.04–0.10) for standard procedures. The fully automated system presented is accurate, fast, environmentally friendly and promising for multiple analytical applications.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.