Abstract

A multiple detector system for γ - γ coincidence and angular correlation measurements is described. The system consists of four large coaxial Ge detectors set at fixed positions, enabling the measurement of γ - γ coincidences between any pair. Thus, six distinct angles are measured simultaneously. The performance of the system was studied with a 152Eu source. An application for γ - γ correlation studies in 146Ce is described. Systematic errors, if present, are shown to be less than 0.03 and 0.05 for A 2 and A 4, respectively. The extension of this system to more than four detectors is straightforward.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.