Abstract

Abstract Techniques for improving the relative accuracy of longwave radiation measurements by a set of pyrgeometers [the Eppley Laboratory Precision Infrared Radiometer (Model PIR)] are presented using 10 PIRs from the 1999 Cooperative Atmosphere–Surface Exchange Study (CASES-99). The least squares–based optimization technique uses a field intercomparison (i.e., a time period during which all the PIRs were upward looking and set up side by side) to determine a set of optimization coefficients for each PIR. For the 10 CASES-99 PIRs, the optimization technique improved the standard deviation of the difference of downwelling irradiance between the PIRs from ±0.75 to ±0.4 W m−2 (for nighttime data). In addition to presenting the optimization method, various PIR data quality checks are outlined and applied to the PIR data. Based on these quality checks, the measured case and dome temperatures of the CASES-99 PIRs were all reasonable. Using the 10 CASES-99 PIRs, simple estimates of the average nighttime net rad...

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.