Abstract

This study used a standard commercial electronic portal imaging device (EPID) area detector attached to an isocentric linear accelerator and the Feldkamp algorithm to produce cone beam tomographic reconstructions. The EPID has a active area of , and can record 12-bit images using two monitor units (MU), with a resolution of FWHM. Since the EPID was not large enough to record the full patient projection at about 1.5 geometric magnification, it was necessary to offset the detector to collect half-cone projections. Corrections are required to convert pixel values into units of exit dose and to realign the projections to overcome the support arm sag. With a geometric magnification of 1.5 the sensitive volume is a cylinder of radius 21 cm and length 17 cm. Unfortunately, the patient couch contains metal bed support rails that lie just outside this cylinder, and produce streak artefacts in the reconstruction. Using 90 views the system delivers a central dose of 90 cGy, and has a density resolution of 4%.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.