Abstract

Abstract. A commercially available dry chemiluminescence (CI) instrument for fast and precise measurement of ozone (O3) is specified. The sensitivity is ~9000 counts s−1 per ppbv of ozone. Its precision is entirely determined by the number of photons reaching the detector (being a photomultiplier), i.e. is quantum-noise limited. The relative precision (ΔO3/O3 in %) thus follows Poisson statistics and scales with the square root of the measurement frequency f and with the inverse O3 mixing ratio: ΔO3/O3 ∝ f0.5 · O3−0.5. At typical O3 mixing ratios between 10 and 100 ppbv (and 1 bar), the precision is 0.3–1.0% at f = 10 Hz. The maximum measurement frequency is 50 Hz. The mechanical and electronic set-up as well as the instrument performance is described. Recommendations on the adequate inlet tube configuration (inlet tube length, sampling flow) and on the way of calibration at stationary ground-based platforms and onboard aircraft are given.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.