Abstract

The ability to visualize intraluminal surface of peritoneal dialysis (PD) catheter and peritoneal cavity could allow elucidation of the cases of outflow problems, and provide information on changes to the peritoneal membrane leading to encapsulating peritoneal sclerosis. A non-invasive examination that allows those monitoring in need is desirable. We have developed a disposable ultra-fine endoscope that can be inserted into the lumen of the existing PD catheter, allowing observation of the luminal side of the catheter and peritoneal cavity from the tip of the PD catheter, with minimum invasion in practice. In a pre-clinical study in pigs and a clinical study in 10 PD patients, the device provided detailed images, enabling safe, easy observation of the intraluminal side of the entire catheter, and of the morphology and status of the peritoneal surface in the abdominal cavity under dwelling PD solution. Since this device can be used repeatedly during PD therapy, clinical application of this device could contribute to improved management of clinical issues in current PD therapy, positioning PD as a safer, more reliable treatment modality for end-stage renal disease.

Highlights

  • The ability to visualize intraluminal surface of peritoneal dialysis (PD) catheter and peritoneal cavity could allow elucidation of the cases of outflow problems, and provide information on changes to the peritoneal membrane leading to encapsulating peritoneal sclerosis

  • In the pre-clinical study of two pigs, we examined the potential risk of peritoneal injury through maneuvering the endoscope as follows; the endoscope was projected from the tip of catheter and strongly scratched the surface of parietal peritoneum under the condition with or without PD solution

  • As shown in the fluoroscopy video of the pre-clinical study (S1), the free part of the endoscope which is projected from the tip of PD catheter may move momentarily

Read more

Summary

Introduction

The ability to visualize intraluminal surface of peritoneal dialysis (PD) catheter and peritoneal cavity could allow elucidation of the cases of outflow problems, and provide information on changes to the peritoneal membrane leading to encapsulating peritoneal sclerosis. In a pre-clinical study in pigs and a clinical study in 10 PD patients, the device provided detailed images, enabling safe, easy observation of the intraluminal side of the entire catheter, and of the morphology and status of the peritoneal surface in the abdominal cavity under dwelling PD solution. Since this device can be used repeatedly during PD therapy, clinical application of this device could contribute to improved management of clinical issues in current PD therapy, positioning PD as a safer, more reliable treatment modality for end-stage renal disease. Non-invasive examination of the intraluminal catheter and peritoneal membrane during the course of PD treatment could have enormous clinical benefit

Methods
Results
Conclusion
Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.