Abstract

The measurement arrangement, which complies with the requirements of the US Food and Drug Administration, consists of a positioning system with a full range of degrees-of-freedom and a digital oscilloscope, both under complete computer control. The acoustic pressure-time waveform is recorded using membrane-type and needle-type polyvinylidene fluoride (PVDF) hydrophone probes. The overall bandwidth of the system depends on the hydrophone probe used and can range up to 100 MHz. A complete description of the system and the measurement procedures is given, along with a brief discussion of the various factors which affect measurement uncertainty. The largest overall uncertainty of the same associated with acoustic intensity measurements was determined to be no greater than 20% for I(sppa) and 25% for I(spta) (spatial peak pulse average intensity and spatial-peak temporal-average intensity, respectively). Other applications of the system include transducer characterization and research work in ultrasound dosimetry and bioeffects.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.