Abstract

Management of the difficult airway is one of the major challenges that anaesthetists face. The flexible fibreoptic scope is widely available but its use requires a level of skill, training and continued practice that is not universally found in all anaesthetists, particularly trainees. The Seeing Optical Stylet is a new, semirigid fibreoptic stylet 'scope. We compared the Seeing Optical Stylet with a gum elastic bougie in a simulated Cormack and Lehane Grade 3 laryngoscopy in a manikin. Forty-four anaesthetists were timed while intubating the manikin's trachea with both devices. The mean (SD) time taken with the Seeing Optical Stylet was 20.8 (9.3) s and with the bougie 30 (19.8) s (p = 0.001). Oesophageal intubation occurred six times with the bougie but did not occur with the Seeing Optical Stylet (p = 0.011). We conclude that the Seeing Optical Stylet may be superior to the bougie in difficult tracheal intubation. We feel that the results of this manikin trial are sufficiently encouraging to proceed to a clinical trial in patients.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.