Abstract

To determine the response to small repeated variations in X-ray exposure of storage phosphor (SP) plates compared with Ektaspeed Plus dental film. Exposure of SP plates and Ektaspeed Plus dental film was varied within a small range [8.62 microC kg-1 to 9.03 microC kg-1, (5%)]. Radiographs were digitized and SP plates read with a laser-based scanner. Nested analysis of variance, analysis of covariance, and regression analysis were used to test the relationship between exposure and the resulting signal. The relationship between exposure and signal for digital images of film radiographs was not significant whereas the corresponding relationship for SP images was highly significant (P < 0.001, R2 = 0.44). Over the narrow range of X-ray exposures studied there was a significant direct linear relationship between exposure and digital gray-scale values for SP images but not for digital images of Ektaspeed Plus films.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.