Abstract

A number of finite element analyses (FEAs) for the dental implant were performed without regard for preload and with all interfaces ‘fixed-bonded’. The purpose of this study was comparing the stress distributions between the conventional FEA model with all contacting interfaces ‘fixed-bonded’ (bonded model) and the model with the interfaces of the components in ‘contact’ with friction simulated as a preloaded implant (contact model). We further verified the accuracy of the result of the FEA using model experiment. In the contact model, the stress was more widely distributed than in the bonded model. From the model study, the preload induced by screw tightening generated strain at the peri-implant bone, even before the application of external force. As a result, the bonded model could not reproduce the mechanical phenomena, whereas the contact model is considered to be appropriate for analysing mechanical problems.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.