Abstract

National Institute of Metrology, China (NIM) and National Physical Laboratory, UK (NPL) blackbody cavities based on the ITS-90 freezing points of In, Sn, Zn and Al have been compared by radiation thermometry. The average differences between In, Sn and Zn, measured by NPL InGaAs radiation thermometer, are 5.3, 2.2 and −7.3 mK, respectively, and the average differences between Sn, Zn and Al, measured by NIM InGaAs radiation thermometer, are 1.4, −5.8 and 13.7 mK, respectively. These differences are well within the uncertainties of the comparison. The NPL blackbody reference sources from 156 °C to 1000 °C were verified using both the NPL and the NIM InGaAs thermometers calibrated at the fixed points. The differences between the radiance temperature of the sources and the temperature measured by the contact sensor within the source are generally within 0.1 K.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.