Abstract

This paper compares a new film-screen system (FSS) called INSIGHT Skeletal Imaging System with the previously used Lanex/T-MAT G FSS. Using a Bronder phantom, measurements were made of dose, resolution and contrast. 135 skeletal phantom images were assessed in order of quality by six observers. Comparable high resolution film-screen combinations (FSC) showed similar geometric resolution. Comparing high intensifying screens, the new INSIGHT Skeletal Regular FSS showed better resolution than the Lanex medium FSC. Dose reduction for the INSIGHT Skeletal Imaging FSS was 29-56%. The new FSS showed image quality similar to high resolution screens but was significantly better when using high intensifying screens. The new INSIGHT Skeletal Imaging System can replace the Lanex/T-MAT G FFS by retaining quality but reducing radiation dose by 29-56%. Using the new high intensifying FSC, images showed in addition improved film quality.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.