Abstract

We present a compact design for a velocity-map imaging spectrometer for energetic electrons and ions. The standard geometry by Eppink and Parker [Rev. Sci. Instrum. 68, 3477 (1997)] is augmented by just two extended electrodes so as to realize an additional einzel lens. In this way, for a maximum electrode voltage of 7 kV, we experimentally demonstrate imaging of electrons with energies up to 65 eV. Simulations show that energy acceptances ≲270 and ≲1200 eV with an energy resolution ΔE∕E≲5% are achievable for electrode voltages ≤20 kV when using diameters of the position-sensitive detector of 42 and 78 mm, respectively.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.