Abstract

This work is concerned with the design of a Cavitation Susceptibility Meter based on the use of a venturi tube for the measurement of the active cavitation nuclei concentration in water samples as a function of the applied tension. The operation of the Cavitation Susceptibility Meter is analyzed and the main considerations leading to the proposed design are illustrated and critically discussed. The results of this analysis indicate that the operational range is mainly limited by nuclei interference, flow separation and saturation (choking), and suggest to develop a Cavitation Susceptibility Meter where: (a) the flow possesses a laminar potential core throughout the venturi throat section in all operational conditions; (b) the pressure at the venturi throat is determined from the upstream pressure and the local flow velocity; (c) the detection of cavitation and the measurement of the flow velocity are carried out optically by means of a Laser Doppler Velocimeter; (d) a custom-made electronic Signal Processor incorporating a frequency counter is used for real time data generation and temporary storage; (e) a computerized system performs the final acquisition and reduction of the data.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.