Abstract

In the measurement system based on four-quadrant detector (4-QD), the non-uniform response and the noise interference can significantly reduce the accuracy of the system. In this paper, the angle measurement principle based on 4-QD is briefly introduced, after that the non-uniform characteristics and noise interference are modeled and simulated respectively. On this basis, a calibration method that can calibrate the noise and non-uniformity of the measurement system is designed, and the corresponding calculation formula is given. Finally, a calibration system is designed and built according to the new calibration method, and multigroup experiment are designed and carried out. From that, the correctness and practicability of the calibration method are verified by the experimental results: compared with the traditional calibration method, the accuracy of the system after calibration is improved by more than 50%, and the noise robustness is better.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.