Abstract
A broadband power meter calibration system based on a newly constructed coaxial calorimeter has been developed in the frequency region 10 MHz-40 GHz. The RF power is measured as the difference of DC power supplied to the calorimeter built-in heater in the RF load, when RF is turned off and on, holding an isothermal control between the RF load and a temperature reference. To minimize the error due to the adiabatic coaxial waveguide, we devised a new method utilizing its output port as a test port. The evaluations showed a calibration uncertainty of (0.28-2.2)% expressed by one standard deviation at the 1 mW level in the full band.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: IEEE Transactions on Instrumentation and Measurement
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.