Abstract
A beat-modulated 118.8-μ m CH3OH laser interferometer system has been constructed and successfully operated for the electron density measurement of JIPP T-II plasma. A Michelson-type far-infrared (FIR) laser interferometer system with a corner-cube reflector has been adopted in order to compensate the mechanical vibration component of the system by means of a 632.8-nm He-Ne laser interferometer. The corner-cube reflector is directly attached to the JIPP T-II vacuum vessel because of simplicity of the interferometer system. It has been observed that the vibration component of the FIR laser interferometer is compensated completely and the electron density by the system agrees closely with the one by a 2-mm microwave interferometer system. The signal-to-noise ratios of the probe- and reference signals are 50 and 250, respectively. The minimum detectable line-averaged density of present system is 8 × 1011 cm-3 (that is, 1/40 fringes).
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.