Abstract

In this paper, the Ansoft high-frequency structure simulation (HFSS) was adapted to investigate the unloaded quality-factor value of the radio-frequency (RF) receiving resonator. This paper focused on the materials aspects, and a comparison between electromagnetic stimulations and measurement was conducted. A 40 mm in diameter Bi2Sr2Ca2Cu3Ox (Bi-2223) tape high-Tc superconducting RF resonator in 3 T was stimulated and built. The simulation models were established according to the experimental dimension of the Bi-2223 surface resonator. Measurements show that the Bi-2223 surface resonator at 77 K provides a gain of 3.84-fold signal-to-noise ratio on phantom images over that of the homemade copper resonator at 300 K. Measuring results were in accordance with predicted ones, and the difference between the predicted SNR gains and measured SNR gains is 1%. This paper suggests that using a Bi-2223 surface resonator at 77 K could be more useful for magnetic resonance imaging coils than a copper surface resonator at 300 K.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.