Abstract

This brief presents an energy-efficient 10-bit accuracy with 20-kS/s successive approximation register analog-to-digital converter for portable pulse oximeter. A data-dependent capacitor reset (DDCR) switching scheme for the capacitive digital-to-analog converter (CDAC) to reduce the average switching energy and the number of unit capacitors is proposed and implemented. Compared with the conventional capacitor switching scheme for CDACs, the proposed DDCR switching scheme reduces the average switching energy and the total number of unit capacitors by 97% and 75%, respectively. We achieved a signal-to-noise-and-distortion ratio of 56.5 dB and a spurious-free dynamic range of 64.7 dBc at the Nyquist input frequency. The measured peak differential and integral nonlinearities are 0.44 and 0.58 least significant bit, respectively. The figure of merit is 9.1 fJ/conversion-step. The prototype, fabricated in the 0.11- $\mu\mbox{m}$ CMOS process, occupies 0.033 mm2.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.