Abstract
Quality control testing of radiographic and fluoroscopic imaging systems requires positioning of test objects in the x‐ray beam in a precise and repeatable fashion. In this work we present several three‐dimensional (3D) printed testing aids that improve efficiency, accuracy, and repeatability of quality control testing. We also present a new device for determining the location of the perpendicular ray in radiographic systems. These devices were designed in an open source software program (OpenScad, http://www.openscad.org) and 3D models were saved in .stl format for printing. The models were printed on either a MakerBot Replicator 2 or Replicator Z18 printer (MakerBot Industries, LLC, Brooklyn, NY). The testing aids were printed using polylactic acid (PLA) filament. To investigate the radiographic characteristics of the PLA used, test articles were printed and used to measure the half‐value layer (HVL) thicknesses in mm of PLA and half‐value densities (HVD) in g/cm2 of PLA for two different colors and over a wide range of radiographic beam qualities, using a portable fluoroscopic c‐arm system. HVL thicknesses of clear PLA ranged from approximately 20 mm at 50 kV nominal tube voltage to 27 mm at 120 kV nominal tube voltage. For green PLA, the HVL thickness was 19 mm at 50 kV tube voltage and 25.7 mm at 120 kV tube voltage. The HVD of clear PLA ranged from 2.37 g/cm2 at 50 kV nominal tube voltage to 3.19 g/cm2 at 120 kV nominal tube voltage. For green PLA, the HVD was 2.35 g/cm2 at 50 kV tube voltage and 3.17 g/cm2 at 120 kV tube voltage. The cost of the devices range from under $2 to approximately $20 in materials. The files used to create the models are freely available at https://github.com/Upstate3DLab/3D-Printed-Radiographic-Test-Tools.
Highlights
Quality control (QC) testing of radiographic and fluoroscopic equipment is an important part of quality assurance (QA) programs, ensuring that the imaging equipment is functioning properly and providing maximum benefit to the patient relative to the radiation dose delivered.[1]
Quality control (QC) testing may be categorized into five general areas: (a) Mechanical inspection of the equipment; (b) Beam geometry tests such as light field, collimator, and central ray alignment; (c) Beam quality, tube output, and patient exposure tests; (d) Systems tests such as automatic exposure control, exposure index, and timer accuracy; and (e) Image quality tests such as detector linearity, resolution, and low contrast detectability
The filmstrip holder consists of two parts: A track that is mounted to the image receptor using tape or other adhesive, and the actual film holder that slides in the track for positioning
Summary
Quality control (QC) testing of radiographic and fluoroscopic equipment is an important part of quality assurance (QA) programs, ensuring that the imaging equipment is functioning properly and providing maximum benefit to the patient relative to the radiation dose delivered.[1]. Quality control (QC) testing may be categorized into five general areas: (a) Mechanical inspection of the equipment; (b) Beam geometry tests such as light field, collimator, and central ray alignment; (c) Beam quality, tube output, and patient exposure tests; (d) Systems tests such as automatic exposure control, exposure index, and timer accuracy; and (e) Image quality tests such as detector linearity, resolution, and low contrast detectability. Not all of these tests are independent of each other, for example patient exposures depend on beam quality, tube output, as well as automatic exposure or automatic brightness control performance
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.