Abstract

To analyze the electric field around bipolar resectoscopes, used in urology, in terms of reasons for late complications after a surgical treatment a flexible multielectrode system was developed to measure the 3-D potential distribution. A high spatial resolution is achieved with the least possible individual measurements under the conditions of a quasi-static electric field. A flexible arrangement and positioning of the measuring points in the vertical direction of the experimental environment enable an adjustable spatial resolution and the selection of the region of interest. The existing influence of the multielectrode system on the measuring results is described and a correction method is presented to achieve significant results. Thus, the multielectrode system is usable for a comparative study of bipolar resectoscopes varying in the arrangement of resection and return electrode.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.