Abstract

A technique for removal of retained lens material is described with a three-port 23-gauge vitrectomy system. Removal of the core vitreous is first performed, followed by removal of the cortical vitreous. All vitreous adhesions to the lens are cleared. The cut rate is then decreased to 1,500 cuts per minute, and vacuum increased to 600 mm Hg. The cortical lens material is cleared first, and then the nuclear material is taken with the same vitrectomy probe using the light pipe to assist in crushing the nuclear fragments. With this technique, even large dense nuclear and cortical retained lens material can be removed from the vitreous chamber without the need for a fragmatome.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.