Abstract

A full evaluation of the uncertainty budget for the ytterbium ion optical clock at the National Physical Laboratory (NPL) was performed on the electric octupole (E3) 2S1/2→2F7/2 transition. The total systematic frequency shift was measured with a fractional standard systematic uncertainty of 2.2×10−18 . Furthermore, the absolute frequency of the E3 transition of the 171Yb+ ion was measured between 2019 and 2023 via a link to International Atomic Time (TAI) and against the local caesium fountain NPL-CsF2. The absolute frequencies were measured with fractional standard uncertainties between 3.7×10−16 and 1.1×10−15 , and all were in agreement with the 2021 BIPM recommended frequency.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.