Abstract

The issues in dynamic load cell have been investigated based on the wave propagation model. It was revealed that the output value of the cell varies depending upon dynamic rigidities, i.e. boundary conditions, of the set up position. For the controversy about dynamic calibration method for load cells, it was disclosed by the experiments that static method of calibration is satisfactory for the column type load cells which usually employed in the measurement of dynamic loads.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.