Abstract

We have developed an analytical method which enables highly precise and rapid quantitative analysis of ultra low sulfur contained in steel sample. The developed method is “ultraviolet fluorescence method after combustion”, which combined high frequency induction furnace and the continuous UV fluorescence analyzer of sulfur dioxide. Despite easy operation like the conventional IR method, the newly developed method showed high sensitivity and good precision. The quantitation limit of sulfur in steel was 0.5 mass ppm. In addition, it was shown that this equipment has sufficient stability as a process control analysis apparatus of the iron mill which continues operation without resting.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.